Original Article

Analysis of Drug-Related Impurties by HPLC in Ciprofloxacin Hydrochloride Raw Material

10.4274/tjps.galenos.2021.70594

  • Derouicha MATMOUR
  • Nadjib HAMOUM
  • Khalil Fateh Eddine HASSAM
  • Nassima HAMDI ZIANI
  • Houari TOUMI

Received Date: 27.05.2021 Accepted Date: 02.09.2021 Turk J Pharm Sci 2022;19(3):293-304 PMID: 35775371

Objectives:

In this study, we report the quality control results of drug-related impurity analysis of seven raw materials of ciprofloxacin hydrochloride marketed in Algeria.

Materials and Methods:

According to the European Pharmacopoeia (Eur. Ph.), high-performance liquid chromatography (HPLC) was used to analyze (B, C, D and E) impurities, while thin layer chromatography (TLC) used to control impurity A.

Results:

HPLC analysis showed that the C1, C2, C3, C4, and C6 samples have individual contents of specified impurities (B, C, D, E), unspecified and the total of all present impurities conform to norms. The C5 sample contains a very high content (0.579%) of impurity C, which is a photodegradation product and the impurities total (0.625%) exceeding limit, while C7 sample has a slightly higher content (0.118%) of unspecified impurity. The control solution of impurity A was not migrated in all developed TLC plates, so the system is not compliant, for this reason, an HPLC analysis protocol was developed.

Conclusion:

The results showed that impurity A content conformed in all samples except for the C6 sample, which has equal content to the limit. Therefore, we recommend revising the detecting technique of impurity A by TLC in the Eur. Ph. or replacing it with a more sensitive technique such as HPLC.

Keywords: Drug-related impurities, specified, HPLC, TLC, ciprofloxacin hydrochloride

INTRODUCTION

The identification and quantification of impurities in raw materials is critical to ensure effective and safe treatment. So, impurity control is a key component and a big challenge in the pharmaceutical industry.1,2 Impurities relate to starting materials, by-products, breakdown products or polymorphs. They can appear at active pharmaceutical ingredient (APIs) production level as well as during or after the formulation process. Their concentrations may change upon storage of the product.2,3

Chemical determination of related impurities in APIs is important because a long exposure at low concentrations, can have undesirable side effects or toxicity and/or may interfere with the drug’s activity.3,4 There are no toxicity studies for the majority of impurities, so impurity analysis is a critical step in quality control.5,6 Therefore, specific requirements for impurities are set by the regulatory authorities.6,7

Ciprofloxacin hydrochloride (CPF HCl) (Figure 1) is a synthetic antibiotic that is part of the list of essential drugs established by the World Health Organization (WHO), manufactured by several generic laboratories in Algeria, their high rate of prescription by clinicians thanks to their numerous indications in the different infections (gynecological, urinary, digestive, and respiratory, etc.). CPF HCl has several associated impurities, which are well described and defined in European Pharmacopoeia (Eur. Ph.) 8th edition. The specified impurities are A, B, C, D, and E, which are individually cited and limited by a specific acceptance criterion, while the impurity F is not specified that is present but limited by an overall acceptance criterion.8 According to Eur. Ph., impurities B, C, D, and E are searched by high-performance liquid chromatography (HPLC), while impurity A by thin layer chromatography (TLC) (Table 1).

In this paper, we analyzed and evaluated the drug-related impurities of seven samples of CPF HCl APIs marketed in Algeria using HPLC.


MATERIALS AND METHODS

Seven samples of CPF HCl were collected from pharmaceutical producers located in Algeria.9 They are labeled as follows: C1, C2, C3, C4, C5, C6, and C7 (Table 2).


Research and quantification of impurities B, C, D, and E by HPLC


Standards, reagents, and apparatus

The standard impurities “CPF HCl for identification of SCR peaks (containing impurties B, C, D, and E)” were purchased from Eur. Ph. (Strasbourg, France). Acetonitrile (HPLC grade),  triethylamine and phosphoric acid were produced by Sigma-Aldrich.8

An HPLC-ultraviolet (UV) device (Thermo Scientific Dionex UltiMate 3000 Rapid Separation LC systems) equipped with an automatic injector and UV detector.


Analysis protocol

Mobile phase: Thirteen volumes of acetonitrile were mixed with 87 volumes of phosphoric acid at 2.45 g/L.

Test solution: 25 mg of CPF HCl of each sample was dissolved in 50 mL of mobile phase.

Control solution (c): 1 mL of the test solution was diluted in 500 mL of mobile phase.

Control solution (b): 2.5 mg of CPF HCl for identification of SCR peaks is dissolved in 5 mL of mobile phase.

Chromatographic conditions: Temperature: 40°C; flow: 1.5 mL/min; injection volume: 50 µL of control solution (b) and (c); detection: 278 nm; column C18: 5 µm, 250 × 4.6 mm.8


Research on impurity A by TLC


Standards, reagents, and apparatus

The standard “impurity A of CPF SCR” was purchased from Eur. Ph. acetonitrile (HPLC grade), ammonia, dichloromethane, and methanol were obtained from Sigma-Aldrich. Silica gel plate F254 for TLC, chromatography tank, and UV lamp at 254 nm were from CAMAG.


Procedure

Test solution: 50 mg of CPF HCl was dissolved in 5 mL water.

Control solution: Impurity A standard (10 mg) is dissolved in a mixture (0.1 mL of diluted ammonia and 90 mL water), completed to 100 mL with water, and 2 mL is diluted in 10 mL of water.

Mobile phase: Acetonitrile, concentrated ammonia, methanol, and methylene chloride (10:20:40:40 V/ V/ V/ V); the deposit volume: 5 µL.

Development: At the bottom of the chromatography tank, a container of 50 mL concentrated ammonia was deposited. The vessel was closed, and the plate was exposed to ammonia vapors for 15 min. The plate was developed on ¾; drying in air and exanimated under UV at 254 nm.


Limits

The sample is compliant, if the impurity A spot is not more intense than main spot of the control solution (0.2%).10


Research and quantification of impurity A by HPLC


Analysis protocol

Mobile phase: Acetonitrile (50 volumes) were mixed with 50 volumes of phosphoric acid at 2.45 g/L.

Standard stock solution: Standard (5 mg) is dissolved in 50 mL of mobile phase.

To determine the maximum absorption of impurity A, the standard solution was scanned in UV over a range of 200 to 400 nm.

The establishment of the calibration curve: Five dilutions were prepared from the standard stock solution (0.1 mg/mL) (Table 3).

Test solution: CPF HCl (50 mg) was dissolved in 5 mL of water.

Chromatographic conditions: Temperature: 25.9°C, flow: 1.5 mL/min, injection volume: 20 µL, detection at 260 nm, column C18: 5 µm, 150 × 4.6 mm.

System compliance: Linearity of the calibration curve with a correlation coefficient greater than 0.990. The symmetry factor of the impurity (A) peak must be between 0.8 and 1.5.

Identification of impurity A: Its retention time.


Results expression

The impurity A content of each sample is expressed by extrapolating its area on the calibration curve: y = a X + b

y: Impurity (A) area, X: Impurity (A) concentration (%)

Calculus formula of impurity A content

Impurity A content (%)=(Peak Area - b)ax50Wieght (mg)

In this study, there was no statistical data analysis.


RESULTS AND DISCUSSION


Research and quantification of impurities B, C, D, and E by HPLC


System compliance

Figures 2 and 3 showed the obtained chromatogram of control solution (b) and typical chromatogram.

These two chromatograms were superimposable and comparable, which enabled us to identify the CPF HCl main peak and impurity E, B, C, and D peaks corresponding.

Retention time (RT) of CPF HCl is 8.962, a value close to that required by Eur Ph that must be at about 9 min. The RT obtained for each impurity (E, B, C, and D) is respectively (3.547 min, 5.977 min, 6.650 min and 11.855 min). All these values are close to those given in the standard chromatogram or calculated from RRT (RT_impurity E: 3.58 min, RT_impurity B: 5.377 min, RT_impurity C: 6.273 min, RT_impurity D: 10.754 min).

The resolution between peaks of impurity B and C is 3, value complies with the required standard (at least 1.3). The symmetry factor of the CPF HCl peak is 1.4, conforming to Eur. Ph. standard (between 0.8 and 1.5). The symmetry factor of peaks belonging to impurities E, B, C, and D (1.16, 1.31, 1.30, and 1.17). All these values were conformed. Therefore, the system compliance is validated.


Sample analysis

Obtained chromatograms of the sample analysis were presented in Figures 4, 5, 6, and 7. Table 4 presents individual contents of (B, C, D, E, unspecified) impurity, and the impurities total.

According to the Eur. Ph. standards, the individual content of impurities B, C, and D must be less than or equal to 0.2%, impurity E, less than or equal to 0.3% and unspecified impurity less than or equal to 0.1%. Any other impurity with individual content less than or equal to 0.05% (exclusion limit) shall not be taken into consideration. The impurities total content shall not exceed 0.5%.

Samples C1, C2, C3, C4, and C6 have individual content of specified impurities B, C, D, and E or unspecified, and the impurities total in the required standards.

C5 sample contains very high content (0.579%) of impurity C compared to the limit, and a total (0.625%) exceeding the norm. This explains that the sample has degraded in impurity C, which is a photodegradation product despite having been well preserved. This result is consistent since the sample was analyzed in date close to its expiry date (August 2016) or it degraded during handling.

C7 sample has individual content of unspecified impurity (known structure such as impurity for unknown structure) equal to 0.118%, slightly higher than the general acceptance criterion and a total in the norm.


Research on impurity A by TLC

Figure 8 shows the TLC plates revelation under UV lamp. The first plate revealed 4 main spots corresponding to test solutions of C1, C2, C3, and C4 samples and no spot of the control solution appeared. The second plate revealed three main spots corresponding to the test solution of C5, C6, and C7 samples and no spot of the control solution appeared. Because of the absence of control migration, a third plate was prepared in which the stock control solution (0.1 mg/mL) was deposited but still has not been migrated.

TLC was re-tested several times, while using

- New reagents to prepare the mobile phase (plate 3);

- New plates silica gel F254 for TLC (plate 3);

- Second control solution prepared from the first vial (plate 4);

- The third control solution was prepared from a second vial of impurity standard (plate 5).

The control was not migrated in all developed TLC plates, so the system is not compliant. For this reason, an HPLC analysis protocol for impurity A was developed.


Research and quantification of impurity A by HPLC


Maximum absorption of the standard solution

Figure 9 shows the absorption spectrum of impurity A in UV.


System compliance

The chromatograms obtained with various standard solutions of the calibration range are shown in Figures 10 and 11.

The maximum absorption of impurity A is 260 nm and its RT is 3.208 min. The symmetry factor of impurity A peak was 1.40, conforming to the norm. The correlation coefficient of the calibration curve is 0.997, which shows that the curve linearity is validated. Therefore, the system is compliant.


Sample analysis

Figures 12 and 13 show the obtained chromatograms of all sample analysis. Table 5 shows the individual contents of impurity A.

According to Eur. Ph., the individual content of impurity A must be less than 0.2%.

Impurity A was not detected in the C2 sample, while C1, C3, C4, C5, and C7 had a content conform but C6 had content equal to the limit.


CONCLUSION

The specified and unspecified impurities (A, B, C, D, and E) was precisely determined in seven samples of CPF_HCl by HPLC. The C1, C2, C3, C4, and C6 samples have individual contents of specified impurities (B, C, D, and E), unspecified and the total of all present impurities conforms to norms. The C5 sample contains very high content of impurity C, which is a photodegradation product and the impurities total exceeding limit, while sample C7 has a slightly higher content of unspecified impurity. Impurity A content is conformed in all samples except for the C6 sample, which has equal content to the limit. According to the detecting technique of impurity A by TLC in the Eur. Ph., the control solution was not migrated, so we recommend revising this method or replacing it with a more sensitive technique such as HPLC.


ACKNOWLEDGEMENTS

I would like to acknowledge Wany Lab laboratory team for their valuable help.

Ethics

Ethics Committee Approval: Not applicable.

Informed Consent: Not applicable.

Peer-review: Externally peer-reviewed.

Authorship Contributions

Concept: D.M., H.T., Design: D.M., Data Collection or Processing: D.M., N.H., K.F.E.H., Analysis or Interpretation: D.M., K.F.E.H., N.H., N.H.Z., Literature Search: D.M., Writing: D.M.

Conflict of Interest: No conflict of interest was declared by the authors.

Financial Disclosure: Therapeutic Chemistry Laboratory, Pharmacy Department, University of Sidi Bel-Abbes, Algeria.

  1. Görög S. Identification and determination of impurities in drugs. Amsterdam: Elsevier Science; 2000. p. 772.
  2. Lee H, Shen S, Grinberg N. Identification and control of impurities for drug substance development using LC/MS and GC/MS. J Liq Chromatogr Relat. 2008;2235-2252.
  3. Görög S. The importance and the challenges of impurity profiling in modern pharmaceutical analysis. Trends Anal Chem. 2006;25:755-757.
  4. Ahuja S, Alsante KM. Handbook of isolation and characterization of impurities in pharmaceuticals. New York: Academic Press-Elsevier; 2003. p. 430.
  5. Chandra MS, Ingale SJ, Paliwal RT, Shivani AK. Advance approaches for the impurity profiling of pharmaceutical. Int J Pharm and Life Sci. 2011;2:955-962.
  6. Farkas SZ, Imre S, Muntean DL, Tero-Vescan A. Analysis of drug related impurties by infrared spectrometry in the class of statins. Farmacia. 2013;6:1091-1101.
  7. Michalska K, Pajchel G, Tyski S. Determination of ciprofloxacin and its impurities by capillary zone electrophoresis. J Chromatogr A. 2004;1051:267-272.
  8. European Directorate for Medicines Quality and Health Care. Ciprofloxacin hydrochloride monograph: related substances. In: European Pharmacopoeia. 8th ed. France: EDQM, 2015. p. 1896-1897.
  9. Ministry of Health, Population and Hospital Reform, Pharmaceutical Products Directorate, Registration Underdirection. Algerian medicines nomenclature, 31th December 2013. Accessed date: January 8, 2014. Available from: http://www.sante.gov.dz/images/pharmacy/nationalnomenclatureofpharmaceuticalproduct
  10. European Directorate for Medicines Quality and Health Care. Ciprofloxacin Hydrochloride Monograph. Impurity A. Thin layer chromatography. In: European Pharmacopoeia. 8th ed. France: EDQM. 2015. p. 1896-1897.