Original Article

Investigation of the Voltammetric Behavior of Methyldopa at a Poly (p-Aminobenzene Sulfonic Acid) Modified Sensor

10.4274/tjps.galenos.2018.44711

  • Gamze ERDOĞDU
  • Şevket Zişan YAĞCI
  • Ebru KUYUMCU SAVAN

Received Date: 22.05.2018 Accepted Date: 31.08.2018 Turk J Pharm Sci 2019;16(4):450-456 PMID: 32454749

Objectives:

The aim was to modify carbon electrodes with (p-aminobenzene sulfonic acid) and use them as a sensor for sensitive and reliable detection of methyldopa (MD) and ascorbic acid.

Materials and Methods:

Electropolymerization was performed by cyclic voltammetry in 0.1 M KCl solution. The modified sensor has a high electrocatalytic effect for oxidation of MD, which appeared in the pH range of 2-11 by differential pulse voltammetry (DPV) techniques.

Results:

For the voltammetric determination of MD, the best results were acquired by DPV in phosphate buffer solution (PBS) (pH 3). The calibration plot of the proposed sensor is linear in two concentration ranges of 1.0-30 and 30.0-300.0 μM. The calibration equations over these ranges are Ipa (μA)=1.21×C (μM)+30.81, R2=0.994 and Ipa (μA)=0.53×C (μM)+53.30, R2=0.9975, respectively. In the sensitivity studies, the limit of quantification and the limit of detection were 10.6 nM and 5.0 nM, respectively. The modified sensor was used for the simultaneous determination of interfering substances such as MD and ascorbic acid in real samples.

Conclusion:

The obtained results revealed that the prepared modified electrode and the proposed method have good sensitivity, repeatability, reproducibility, and stability.

Keywords: Methyldopa, voltammetry, poly (p-aminobenzene sulfonic acid), ascorbic acid, glassy carbon electrode

INTRODUCTION

Methyldopa (MD) is a catecholamine that is known by its chemical name 2-methyl-3-(3,4-dihydroxyphenyl)-DL-alanine (Figure 1), and it is widely used to lower blood pressure. MD is a centrally acting adrenal receptor that reduces high blood pressure and sympathetic tone.1 In adrenergic nerve terminals, it is converted to α-methyl noradrenaline, and its antihypertensive effect seems to be due to this agent stimulating the central adrenoreceptors.2

Various methods like high-performance liquid chromatography with ultraviolet (UV) detection,3 polarography,4 potentiometry,5 UV visible spectrophotometry,6 and flow injection techniques7,8 were reported previously for the determination of MD. However, many of these techniques require expensive equipment and are time consuming. In addition, since these catecholamines are electrochemically active, it is also possible to determine the nature of the molecules that provide neurotransmission by electrochemical methods. Therefore, it is important to detect MD in the presence of ascorbic acid (AA) by a reliable method that has good selectivity and sensitivity.

AA (vitamin C) is a biologically and industrially important substance.9 The coexistence of AA, MD, and other catecholamines with very similar oxidation potentials leads to the response obtained by electrochemical techniques. For this reason, the increased sensitivity and selectivity of the new sensors produced to the MD has long been the subject of research. Using polymer modified electrodes solves this problem. The electrochemical behavior of MD was studied at various polymer electrodes.10,11,12,13,14,15,16,17

However, some disadvantages exist in the previously reported modified electrodes. AA exists as an anion in physiological pH (7.4), whereas MD exists as a cation. There are high electron density sulfo groups and electron-rich N atoms in the structure of p-aminobenzene sulfonic acid (ABSA). For this reason, a negatively charged polymer film is required to eliminate the interference of AA in the determination of MD. The p-ABSA molecule has high electron density sulfo groups and p-ABSA films are negatively charged. Due to the electrostatic repulsion between the negatively charged sulfo groups and the AA anions in the modified sensor, the AA shifts to a more negative potential and the dopaminic acid can be easily separated from it. The p-ABSA modified sensor can show high selectivity against MD.18

In the present study, electroanalytical methods were developed to detect MD in drug samples rapidly, reliably, and sensitively using an electrode modified with poly (p-ABSA). It has been determined that the modified sensor can be utilized for MD determination even in the presence of AA at the same time. Another significant advantage of these techniques over other ones is that they can be applied directly to the analysis of pharmaceutical dosage forms and biological samples without the need for extensive sample preparation, as there is no interaction between the adjuvants and the endogenous substances.

The analytical determination parameters such as the limit of detection (LOD), the limit of quantification (LOQ), and the concentration range were determined, and the amount of MD in the drug tablets and blood serum was found. To test the accuracy of the applied voltammetric method, MD recovery studies were performed in real samples.


MATERIALS AND METHODS


Materials

Alfamet tablets containing 250 mg of MD were kindly supplied by I.E. Ulagay (Turkey). All chemicals were of analytical purity and were procured from Merck (Darmstadt, Germany) or Sigma Chemical Company. Prior to the polymerization the solutions of monomer were held in nitrogen gas atmosphere for about 10 min, and during the electropolymerization the electrochemical cell was covered with nitrogen gas. Voltammetric experiments were carried out in phosphate buffer solution (pH 3.0). MD and AA solutions were freshly prepared before the experiments. All solutions were prepared with ultrapure water.


Instrumentation

In the voltammetry experiments, a BAS (Bioanalytical Systems, Inc.) 100BW electrochemical analyzer was used. This analyzer is connected to a personal computer and the device is controlled and data stored and processed by means of software loaded and running under MS Windows. An electrode system consisting of a Ag/AgCl reference electrode (CHI), a glassy carbon disc working electrode (geometric area: 6.85 mm2, CHI) and a Pt wire coil as auxiliary electrode (CHI) was used.


Modification of poly (p-ABSA) sensor

Before modification, the working glassy carbon electrode (GCE) was cleaned using 0.3 and 0.05 µm Al2O3 slurry on polishing materials. Then the polished GCE was sonicated in 1:1 nitric acid solution for 10 min and washed with ultrapure water. Afterwards, the GCE was electrochemically cleaned by cycling 20 times in the potential range of -0.7 to 1.7 V with a scan rate of 100 mV/s in 0.5 M H2SO4. After that, the electrode was plunged into 0.1 M KCl solution containing 5.0 mM p-ABSA and the modification procedure was performed by cyclic sweeping from -1.5 to 2.5 V for 14 cycles at 50 mV/s. Then the modified sensor was conditioned by cyclic voltammetry in the potential range of -0.5 to 0.5 V with 100 mV/s in pH 3.0 phosphate buffer solution (PBS) and was stored in PBS (pH 3.0).


Preparation of real samples

Tablets of MD with the commercial name Alfamet were prepared. Each tablet contains 250 mg of MD. Five MD tablets were finely powdered using a mortar and pestle and then an appropriate amount of this sample containing a known amount of the active material was weighed and dissolved with double distilled water. The prepared mixture was filtered using filter paper and diluted to appropriate amounts with double distilled water. The serum samples were collected from a research hospital and were sonicated (15 min with 5000 rpm) and then diluted 10 times with double distilled water without any additional pretreatment. Before voltammetric determination, appropriate amounts of the prepared real samples were added to 10 mL of phosphate buffer solution with optimum pH (pH 3.0), followed by transfer to the electrochemical cell for electrochemical measurements. The standard addition method was used to determine MD in the real and spiked samples.


RESULTS


Electropolymerization of p-aminobenzene sulfonic acid

Figure 2 shows the electropolymerization of p-ABSA at the GCE surface. The electropolymerization was performed in 0.1 M KCl solution containing 5.0 mM p-ABSA at a GCE by cyclic voltammetry in the potential range of -1.5 to 2.5 V. In the first cycle, two reduction peaks were obtained at 0.452 V (peak A) and 0.449 V (peak B), which might be related to the reduction of p-ABSA. Again, in the first cycle, an oxidation peak was observed at 0.824 V (peak C). In the next and subsequent cycles, following the continuous scan, broader peaks were monitored, indicating that the polymer film was constantly growing. It could be observed that film growth was faster for the first five cycles than for the other cycles and also the next cycles no longer existed. These findings show that p-ABSA was coated on the GCE surface by electrodeposition. A brown polymer was formed that was properly bonded on the GCE surface.


The effect of film thickness on MD response

The film thickness, which is determined by the number of cycles of electropolymerization, is one of the most important factors determining the polymer film selectivity property. By altering the amount of charge consumed during electropolymerization, it is possible to obtain poly (p-ABSA) films at desired thicknesses. Different film thicknesses were obtained by varying the cycles of the cyclic voltammetry. The selectivity of poly (p-ABSA) sensors prepared in the range of 8-18 cycles to MD and AA was systematically examined. From the DPV results of MD, it was observed that regular and repetitive responses could be obtained at 14 cycle film thickness. This can be seen from Figure 3. Furthermore, the effect of the number of cycles on the electropolymerization was calculated as 64.42%. This value was calculated from the ratio of the highest peak current to the peak current of the first polymer film.


Electrochemical behavior of MD at poly (p-ABSA) modified sensor

The voltammograms achieved by cyclic voltammetry of MD show a reduction wave at a potential of nearly 200 mV and an oxidation peak of nearly 220 mV (Figure 4). The electrochemical oxidation of MD was studied by cyclic voltammetry at the surface of the bare and poly (p-ABSA) modified GCE. The oxidation of MD shows a weak peak on the bare GC at nearly 0.590 V but the experimental results for the modified GCE show a well-defined anodic peak at the peak potential of 0.220 V with respect to the Ag/AgCl reference electrode. The peak current and peak potential values recorded at the GCE electrode were 0.61 µA and 0.590 mV, respectively. However, at the poly (p-ABSA) electrode these values were observed to be 30.48 µA and 0.220 mV, respectively (Figure 4). Consequently, in comparison with the data recorded from the bare GCE electrode, an increase in peak current and a decrease in overpotential of MD were obtained at the modified GC electrode. Therefore, it was assessed as an electrocatalytic effect for the oxidation of MD on the modified surface. It could be observed that the oxidation peak current for the modified electrode significantly increased and it was almost 50 times higher than that for the unmodified electrode. This behavior is due to adsorption of MD on the surface of p-ABSA by interaction of MD functional groups such as NH2, COOH, and OH with carboxyl groups of activated p-ABSA on the surface of the electrode. Thus sensitivity was significantly enhanced due to preconcentration of MD on the active surface of p-ABSA. Moreover, as shown in Figure 4B, the onset potential for MD oxidation at the poly (p-ABSA) electrode is lower than its oxidation at the bare GCE (Figure 4B) because of the catalytic behavior of the modified electrode. However, the potential peak at the bare GCE (0.59 V) is higher than the potential peak at the modified GCE (0.220 V). The effect of scan rate on the oxidation peak current of 0.01 mM MD was studied. With the scan rate increasing, the anodic peak current increased. A good linearity between the square root of scan rate and peak current was obtained in the range of 10-250 mV s-1. The linear regression equation was Ip(µA)= 0.502v1/2–0.899 with correlation coefficient R2=0.998. The correlation coefficient is very close to 1.0, showing that the oxidation process is diffusion controlled. Furthermore, the plot of the logarithm of peak current versus logarithm scan rate has a slope of 0.63, which is almost the theoretical value of 0.56. The equation was logIp(µA)=0.63logv–0.7041 (R2=0.998) on the modified electrode. This indicates a diffusion controlled electron process of MD oxidation at the poly (p-ABSA) modified GCE.


DISCUSSION

The electrostatic interaction between the modified GCE electrode and MD contributed to the enhancement of sensitivity and electroactivity. The oxidation peak of MD at pH 3.0 is irreversible and thus with an increase in peak height the peak potential shifts to lower potential. However, onset potential, which shows the kinetic of the reaction, decreased for the modified GCE compared to the bare GCE and thus sensitivity and selectivity increased because of these effects.


Electrolyte type effect on voltammetric behavior of MD

By selecting an appropriate supporting electrolyte solution, a conductive environment is created between the submerged electrodes.

The choice of supporting electrolyte depends on MD’s resolution, dissociation degree, and nucleophilic character. For this purpose, voltammograms of MD in Na2SO4, PBS (pH 7.0), NaNO3, NaClO4, NaCl, and KCl supporting electrolytes (electrolytes concentration, 0.1 M) were recorded (Figure 5). While a voltammogram was taken at pH 7.0 for PBS, voltammograms were taken at the native pH of the other electrolyte species.


Effect of pH on the peak potential and peak current of MD

The peak current and potential are dependent on the pH of the solution. To find the optimum pH, the influence of pH over the range of 2.0-11.0 on the performance of the sensor was investigated. Experimental results for MD are shown in Figure 6. It was found that the anodic peak current of MD increased with increment of acidity, and reached its maximum value at pH 3.0. Therefore, pH 3.0 was selected as the optimum pH for the determination of MD. Increasing peak current with the increase in acidity showed that the mechanism for oxidation of MD was a proton dependent reaction.

It was observed that as the pH of solution was increased, the oxidation peak potential shifted to negative potential values. The negative shift and the peak potential showed a linear relationship with a slope of -52.4 mV/ pH in the pH range of 2.0-5.0. This slope approximately revealed that the number of protons in the process was equal with the number of electrons transferred in the oxidation reaction of MD.


Determination of MD in the presence of AA

Determination of MD in poly (p-ABSA) was done with differential pulse voltammetry. Differential pulse voltammograms of different concentrations of MD on the poly (p-ABSA) modified GCE are shown in Figure 7. The data of the obtained calibration charts are shown in Table 1. The calibration plot of the proposed sensor is linear in two concentration ranges of 1.0-30.0 and 30.0-300.0 µM.

The calibration equations over theses ranges are Ipa (µA)=1.21×C (µM)+30.81, R2=0.9944, and Ipa (µA)=0.53 C (µM)+53.30, R2=0.9975, respectively. LOD and LOQ were calculated as 5.0 nM and 10.6 nM (S/N=3), respectively. The relative standard deviation (RSD) for MD and 5 repeated measurements was less than 3%.

It is readily seen from Figure 8 that peak currents increase linearly with increasing MD concentration even in the presence of AA. In addition, the MD peak current was unaffected by the increasing AA concentration (Figure 9). Moreover, from the successive runs of the modified electrode in the binary mixture, it was observed that the voltammetric responses were almost invariable. The RSD for MD and 5 repeated measurements was less than 3%. This reflects that the stability of the modified electrode was satisfactory.


Analytical applications

Five Alfamet tablets containing 250 mg of MD in each tablet were directly weighed and powdered in a mortar. The calculated amount of MD corresponding to 100 mM concentration stock solution was weighed and transferred to a 10-mL volumetric flask and the volume was supplemented with ultrapure water. The contents of the flask were subjected to centrifugation at 5000 rpm for 15 min to effect complete dissolution. The prepared mixture was filtered using paper filter and diluted using appropriate amounts of double distilled water. The serum samples were collected from a research hospital and were centrifuged (15 min at 5000 rpm) and then diluted 10 times with double distilled water without any additional pretreatment. Before the voltammetric determination, appropriate amounts of prepared real samples were added to 10 mL of phosphate buffer solution with optimum pH (pH 3.0) and then transferred to the electrochemical cell for electrochemical measurements. The standard addition method was used to determine MD in the real and spiked samples.

The quantity of MD in the tablets was computed from the suitable calibration graphs. Furthermore, the accuracy of the proposed techniques was checked by carrying out recovery studies. Recovery results obtained from the calibration graph can be seen in Table 2. The proposed method was successfully performed on real samples in the presence of interferences.


CONCLUSION

A poly (p-ABSA) modified electrode was applied for electrocatalytic assay of MD. The modified GCE indicated high electrocatalytic activity for MD. The modified GCE provides much sensitivity and selectivity in the assay of MD. Moreover, the modified electrode showed easy regeneration and good repeatability and stability. The modified GCE can be used under the selected conditions (in PBS, pH 3) for the determination of MD. The results show that the proposed method can be easily used in the determination of MD in drug samples and clinical analyses. It has been observed that this method can be used to identify MD in blood serum.


ACKNOWLEDGEMENTS

The Research Fund Unit of İnönü University financially supported this study (Grant no. APYB: 2010/30).

Conflict of Interest: No conflict of interest was declared by the authors.

  1. Hoffman BB, Lefkowitz RJ. In: Gilman AG, Hardman JG, Limbird LE, Molinoff PB, Ruddon RW, eds. The Pharmacological Basis of Therapeutics (9th ed). New York; McGraw-Hill; 1996:1908-1984.
  2. Katzung EG, Masters SS, Trevor AJ. In: Katzung BG, ed. Basic and Clinical Pharmacology (12th ed). New York; McGraw-Hill; 2012:79-96.
  3. Zecevic M, Zivanovic L, Agatonovic SK, Minic D. The use of a response surface methodology on HPLC analysis of methyldopa, amiloride and hydrochlorothiazide in tablets. J Pharm Biomed Anal. 2001;24:1019-1025.
  4. Ballantine J, Woolfson AD. Determination of catecholamines in pharmaceutical preparations by differential pulse polarography at the glassy carbon electrode. Int J Pharmaceutics. 1979;3:239-246.
  5. Badawy SS, Issa YM, Tag-Eldin AS. Potentiometric determination of L-dopa, carbidopa, methyldopa and aspartame using a new trinitrobenzenesulfonate selective electrode. Electroanalysis. 1996;8:1060-1064.
  6. Wahbi AM, Abdine H, Korany M, Abdel-Hay MH. Sodium cobaltinitrite as a reagent for determining some phenolic drugs. J Ass Anal Chem. 1978;61:1113-1117.
  7. Nagaraja P, Vasantha RA, Sunitha KR. A new sensitive and selective spectrophotometric method for the determination of catechol derivatives and its pharmaceutical preparations. J Pharm Biomed Anal. 2001;25:417-424.
  8. Abdulrahman LK, Al-Abachi AM, Al-Qaissy MH. Flow injection-spectrophotometric determination of some catecholamine drugs in pharmaceutical preparations via oxidative coupling reaction with p-toluidine and sodium periodate. Anal Chim Acta. 2005;538:331-335.
  9. Levine M. New concepts in the biology and biochemistry of ascorbic acid. N Engl J Med. 1986;314:892-902.
  10. Kajiya Y, Sugai H, Iwakura C, Yoneyama H. Glucose sensitivity of polypyrrole films containing immobilized glucose oxidase and hydroquinonesulfonate ions. Anal Chem. 1991;63:49-54.
  11. Arrigan DWM, Lowens MJ. Polypyrrole films doped with an electroactive sulfonated chelating reagent: electrochemical characterization and the detection of metal ions. Electroanalysis. 1999;11:647-652.
  12. Gholiv MB, Amiri M. Preparation of polypyrrole/nuclear fast red films on gold electrode and its application on the electrocatalytic determination of methyl-dopa and ascorbic acid. Electroanalysis. 2009;21:2461-2467.
  13. Gao Z, Ivaska A. Electrochemical behaviour of dopamine and ascorbic acid at overoxidized polypyrrole(dodecyl sulphate) film-coated electrodes. Anal Chim Acta. 1993;284:393-404.
  14. Gao Z, Zi M, Chen B. Permeability controllable overoxidised polypyrrole film modified glassy carbon electrodes. Anal Chim Acta. 1994;286:213-218.
  15. Saeed S, Reyhaneh-Sadat S, Zahra K. Sensitive electrochemical sensor for determination of methyldopa based on polypyrrole/carbon nanoparticle composite thin film made by in situ electropolymerization. Electroanalysis. 2011;23:2248.
  16. Behzad  R, Neda A, Ali AE. Adsorptive stripping voltammetry determination of methyldopa on the surface of a carboxylated multiwall carbon nanotubes modified glassy carbon electrode in biological and pharmaceutical samples. Colloids and Surfaces B Biointerfaces. 2013;109:253-258.
  17. Seyed ZM, Hadi B, Malihe J, Ali A. Nanomolar determination of methyldopa in the presence of large amounts of hydrochlorothiazide using a carbon paste electrode modified with graphene oxide nanosheets and 3-(4′-amino-3′-hydroxy-biphenyl-4-yl)-acrylic acid. Electroanalysis. 2015;27:2421-2430.
  18. Diaz A, Bargon J, Handbook of Conducting Polymers, Skotheim, TA, ed, New York; Marcel Dekker Inc. 1986:1:81-82.