DETERMINATION OF NAPROXEN SODIUM FROM POLY(LACTIDE-CO-GLYCOLIDE) CORNEAL SCAFFOLDS
PDF
Cite
Share
Request
Original Article
P: 57-68
April 2010

DETERMINATION OF NAPROXEN SODIUM FROM POLY(LACTIDE-CO-GLYCOLIDE) CORNEAL SCAFFOLDS

Turk J Pharm Sci 2010;7(1):57-68
1. Hacettepe University, Faculty Of Pharmacy, Department Of Pharmaceutical Technology, 06100, Ankara-Turkey
2. Hacettepe University, Faculty Of Pharmacy, Department Of Analytical Chemistry, 06100, Ankara-Turkey
No information available.
No information available
Received Date: 22.10.2009
Accepted Date: 25.02.2010
PDF
Cite
Share
Request

ABSTRACT

A simple, very specific, and reliable HPLC method has been developed and validated for the determination of naproxen sodium (NS) in poly(lactide-co-glycolide) (PLGA) (50:50) scaffolds. Efficient chromatographic separation was achieved on a reverse phase column (ODS2 C18 200x4.6mm) with the mobile phase consisted of methanol and acetate buffer (pH 5.1) (55:45 v/v) at a flow rate of 1 mL.min-1 by using fluorescence detector at 254 nm ex/ 352 nm em wavelength. The retention time was about 9.3 min. Calibration curve was linear over the concentration range of 0.08-75 ug.mL-1. The intra- and inter-day precision relative standard deviation was 3.71 % or less, and the accuracy was within 2.96 % of the nominal concentration. The developed HPLC method was successfully applied to quantitate NS in PLGA (50:50) scaffolds.

Keywords:
Naproksen sodyum, PLGA, HPLC, Validasyon